Factory Calibration

Recommended Recalibration Frequency: No regular factory recalibration required.

What's included with LI-600/N recalibration?

A factory recalibration includes:

Calibration

  • Calibrate flow meter across range (10 flow rates across flow range). Performance is verified after calibration.
  • Calibrate RH sensors (dry air and up to 6 dew points at 3 ambient temperatures). Performance is verified after calibration.
  • Infrared leaf temperature sensor (calibration is derived from 5 temperature measurements of a simulated leaf in 3 ambient air temperatures that span the operating temperature range). Performance is verified after calibration.

Tests and Maintenance

  • Check filters and replace if needed.
  • Check gaskets and replace if needed.
  • Test battery.
  • Verify functionality.
  • Other repairs and maintenance as needed. Customer approval is required for additional repairs.

To request factory service or recalibration for an instrument:

Service will be performed at our US office in Lincoln, NE USA, European office (LI-COR GmbH), or by your local distributor.