Specifications

  • Dew Point Range: 0 to 50 °C (limited to 35 °C below the cooling water temperature)
  • Accuracy: ± 0.2 °C (0-50 °C)
  • Stability: <0.02 °C per day at 25 °C typical; <0.04 °C per day at 25 °C maximum
  • Noise Level: 0.01 °C peak-to-peak
  • Repeatability: 0.01 °C
  • Response Time: Typically 15 seconds per °C when changing from ambient to a higher dew point; 30 seconds per °C for dew points lower than ambient
  • Temperature Sensor: Platinum resistance temperature detector (RTD)
  • Flow:
    • Flow Rate: Adjustable; typically 2.0 liters per minute
    • Flow Meter Type: Dwyer series RMA. 2.5 liters per minute full scale
    • Flow Meter Accuracy: ± 4% of full scale reading
    • Flow Outputs: Two rapid connect hose fittings for 4 mm ID by 6 mm OD plastic tubing. Accepts 1/4" OD tubing with 1/8 to 3/16" ID
    • Maximum Input Flow: 2 liters per minute. Contact LI-COR about using higher input flow rates
  • Analog Output: 0 to 5 V, 100 mV / °C
  • Command Input: 0 to 5 V, 100 mV / °C
  • Display: 4 1/2 digit LCD for displaying dew point set temperature (°C), actual dew point temperature (°C), or battery voltage
  • Display Resolution: 0.01 °C
  • Operating Temperature: 0 to 50 °C, 0 to 100% RH
  • Power Requirements: 10.5 to 16 VDC, 5.5 amps max current draw; or 108-126/216-252 VAC with 610-01 AC Module
  • Dimensions: 23.5 H x 21 W x 28.5 D cm (9 H x 8.1 W x 11" D)
  • Weight: 7.86 kg (17.4 lbs.)